PT0E-0730 RIKEN PORTABLE COMBINATION GAS DETECTOR MODEL RX−516 RX−517 OPERATION MANUAL MARINE USE Precaution in operation
PT0E-0730 93−3.Start-up and end-up When press power switch for 3 sec, the power gets on and this shall start up sequencing the program of the volta
PT0E-0730 103−4.Flow monitor check When the power gets ON, the pump shall start operation check that suction works properly by filter tube with flo
PT0E-0730 11<FLOW CHART OF ZERO POINT ADJ> ZERO MODE DISPLAYHOLD KEY ADJ ZERO OK 《Zero set OK》 ZERO SET OK FAIL ZERO ・Press ZERO key for 3 sec. Pre
PT0E-0730 123−6.Operational procedure Approach the sampling hose to the sampling site for measurement and leave sampling gas sucked. ・Take care t
PT0E-0730 133−7.Peak SW value mode When press PEAK SW during operation, the peak valve from power on to the current point (HC, H2S = MAX, O2=MIN) s
PT0E-0730 14 3−9.Operation finished (1)Treatment after operation Wind up sampling hose so treat it can not be twisted and bent. Supply fresh air
PT0E-0730 153−10.Practical application (1) Tank inside of crude oil tanker By use of sampling hose with plummet, it is available to monitor the gas
PT0E-0730 16《Interference measures in inert gas》 In the component of inert gas, a lot of interference gas may be present. When it should be present s
PT0E-0730 174−3.Trouble display (1)Low flow alarm ・When low flow trouble triggers, “LOW FLOW” is displayed and pump stops working. ・Buzzer sounds con
PT0E-0730 18 (3)System trouble alarm ・When the program is troubled, “SYSTEM FAIL” will be shown. ・Buzzer will sound continuously. ・After “SYSTEM FAIL
PT0E-0730 1 In the beginning It is of our great pleasure to purchase Riken portable combination gas detector model RX-516/RX-517 this time. This
PT0E-0730 195.Each function 5−1.Operation mode control Operation mode HOLD KEY D A T E T I M E ①1st alarm display ②2nd alarm Day/time display Data l
PT0E-0730 205−1−1.Alarm setting 1st alarm display PEAK key M ‐ P A S S W O R D ・First, 1st alarm point shall be shown on the display. ・Change 『+』
PT0E-0730 215−1−2.Date/time setting 5−2.LCD back light LCD back light will be on in the following ・Initial clear : Light on ・Operation on
PT0E-0730 225−3.Data logger Data logger function has following 5 kinds. (1)Interval trend The change of measuring gas shall be recorded until the po
PT0E-0730 236.Maintenance check For this correct operation, follow to the following procedure. 6−1.Battery replacement When replace the
PT0E-0730 246−2.Gas calibration (Span adjustment) ZERO key PUMP key ・ Hold and pressing ZERO and PUMP key similtaneously in the operation mode
PT0E-0730 25 《Handling of sampling bag and canned gas》 ① Sample the cal gas into sampling bag from canned gas. ② Sampling bag with cal g
PT0E-0730 26《Cal gas make-up by H2S gas checker CK-82》 ① Take out 1 pce. ampoule from accessory ampoule pack. ② Check that the rubber caps are fix
PT0E-0730 276−3.Filter replacement For filter (cotton) inside of filter with flow monitor, check it before use and when it gets dirty, replace it.
PT0E-0730 286−6.The replacement of parts The following parts have a life and each, regular replacement shall be recommended. When replace, contac
PT0E-0730 2 C O N T E N T S 1.Caution in operation ・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 3 2.Name of each part・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・・ 4 3.Operatio
PT0E-0730 296−7.Storage or treatment not to use for a long period The instrument shall be kept in the following condition. Temp :10∼30℃ Humidity :30∼
PT0E-0730 308.Trouble shooting This trouble shooting does not mention all the trouble causes. The following is mentioned briefly by picking up the
PT0E-0730 319.Definition of words Vol% Gas density is expressed by the unit of a part. ppm Gas density is expressed by the unit of a part per.
PT0E-0730 3210.Specifications Refer to the specifications for separate catalog overleaf. Model RX-516 (HC/O2/H2S 0-100.0ppm) , RX-517 (HC/O2/H2S 0
PT0E-0730 3311.Detection principle 11−1.NDIR method (Non-Dispersive Infrared) The model is based on NDIR method (Non-Dispersive Infrared) and this st
PT0E-0730 34 11−3.ELECTRO-CHEMICAL CELL method This gas detector applies an electrochemical sensor. The sample gas is electrolyzed by the electrolyze
PT0E-0730 31.Caution in operation To maintain the function for safety, follow the following instruction. Warning
PT0E-0730 42.Name of each part ①POWER/ENTER SW This is used for power ON/OFF and entering the figure. ②PUMP/− SW This is used when make the pump
PT0E-0730 5《Display of RX-516/517》 Hydro carbon 「HC」 RX-516/RX-517 Hydrogen sulfide(Low)「H2S」 Oxygen「OXY」RX-517 Hydrogen sulfide(High)「H2S」*Caut
PT0E-0730 6【Standard accessories】 ①Carrying case with shoulder strap ②Filter tube with flow monitor and junction tube ③G
PT0E-0730 7 【Optional accessories】 ①Gas sampling hose 30m with plummet ②Gas sampling hose 30m with float type sampling head
PT0E-0730 83.Operation 3−1.Check item before operation RX-516 or 517 instrument: ・Check that there is no damage for display unit etc. Filter tube wit