Rki-instruments SM-2009U User Manual

Browse online or download User Manual for Sensors Rki-instruments SM-2009U. RKI Instruments SM-2009U User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - Calibration Station

www.rkiinstruments.com Operator’s ManualSM-2009U Single ModuleCalibration StationPart Number: 71-0164RKRevision: AReleased: 8/30/11

Page 2 - Warranty

6 • Air Filter & Sample TubingWARNING: Do not use an exhaust tube that is longer than 10 feet. Using an exhaust tube longer than 10 feet will cau

Page 3 - Table of Contents

Instrument Cradle • 7Instrument CradleThe instrument cradle is a recessed area on the top of the SM-2009 that is designed to accept the GX-2009. Inse

Page 4

8 • Back PanelBack PanelThe back panel includes the power jack and sample fittings.Power JackThe power jack is located in the lower left corner of th

Page 5 - About the SM-2009U

Control Panel • 9Control PanelThe control panel is used to setup and operate the calibration station. It is located at the front of the calibration s

Page 6 - System Requirements

10 • Control PanelTable 2: Control Button FunctionsControl ButtonControl Button Function(s)Control Button LED Function(s)BUMP ▼• Initiates a bump tes

Page 7 - Specifications

USB Port • 11USB PortA USB port is located on the right front of the calibration station below the COPY and POWER buttons. The port can be used to sa

Page 8 - About this Manual

12 • OverviewChapter 3: Preparing to Use the SM-2009UOverviewThere are four tasks that must be completed before you can begin to use the SM-2009U: ha

Page 9 - AC Adapter

Setting the Bump Test and Calibration Parameters • 13CAUTION: The maximum recommended length for the exhaust tube is 10 feet. Do not use more than 10

Page 10 - Particle Filter for Air Inlet

14 • Setting the Bump Test and Calibration ParametersAir Sample Time (Air)The air sample time can be set separately for bump testing and calibration.

Page 11 - Instrument Cradle

Setting the Bump Test and Calibration Parameters • 15calibration gas through the CAL. GAS fitting on the back of the station during a bump test.Bump

Page 12 - Back Panel

WarrantyWarrantyRKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials and workmanship, and performance f

Page 13 - Control Panel

16 • Setting the Bump Test and Calibration Parameters 5. Install a GX-2009 in the instrument cradle. The CHARGE LED will begin to blink yellow. See “

Page 14 - 10 • Control Panel

Setting the Bump Test and Calibration Parameters • 17NOTE: You can still use the SM-2009 with an instrument for bump testing and calibration if this

Page 15 - USB Port

18 • Setting the Bump Test and Calibration Parameters 2. Press and release the EDIT ENTER button. The GX-2009 will display the following screen. 3. P

Page 16 - Hardware Assembly

Setting the Bump Test and Calibration Parameters • 19 5. Use the BUMP ▼ and CAL. ▲ buttons to scroll through the parameters until the desired paramet

Page 17 - Parameters

20 • Setting the Bump Test and Calibration Parameters 7. Use the BUMP ▼ and CAL. ▲ buttons to set the parameter to the desired value, then press and

Page 18

Setting the Bump Test and Calibration Parameters • 21To continue, press and release the EDIT ENTER button. The first parameter, air sample time, will

Page 19

22 • Connecting Calibration Gas 7. Use the BUMP ▼ and CAL. ▲ buttons to set the parameter to the desired value, then press and release the EDIT ENTER

Page 20 - 50 25

Connecting Calibration Gas • 23To connect calibration gas to the SM-2009U do the following: 1. If the area around the calibration station is not cons

Page 21 - TRANSMIT

24 • Installing the Single Module Data Viewer SoftwareInstalling the Single Module Data Viewer Software 1. Launch Windows®. 2. Exit from all applicat

Page 22 - 3 0

Overview • 25Chapter 4: Using the SM-2009UOverviewWhen you have completed the tasks in "Chapter 3: Preparing to Use the SM-2009U", you are

Page 23

Table of ContentsTable of ContentsChapter 1: Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1Overview . .

Page 24 - BUMP/CAL

26 • Bump Testing Instrumentsbegin to blink yellow. If the batteries are fully charged, the CHARGE LED will become solid green after about 5 minutes.

Page 25 - E GAS - IN

Bump Testing Instruments • 27connecting to the SM-2009U.The failure screen will remain on the display while the unit is connected to the SM-2009U.NOT

Page 26 - Connecting Calibration Gas

28 • Bump Testing Instruments 10. Press and hold the BUMP ▼ button for at least one second, then release it. The bump test begins. During the bump te

Page 27

Bump Testing Instruments • 29apply gas and perform a calibration. 16. The BUMP LED will stop blinking and be on steadily green if the bump test passe

Page 28 - Viewer Software

30 • Bump Testing Instruments 19. The results of the bump test or bump test and calibration will be stored in the SM-2009U’s memory and will be avail

Page 29 - Bump Testing Instruments

Calibrating Instruments • 31Calibrating InstrumentsWhen a calibration is performed, the calibration station performs a fresh air adjustment on an ins

Page 30

32 • Calibrating InstrumentsNOTE: The screen shown above applies to a 4-channel GX-2009. If your GX-2009 has less than 4 channels, the inactive chann

Page 31

Calibrating Instruments • 33• When Cal Limit Display is set to On and Cal. Limit Check is set to Confirm to use, if the GX-2009 is due for calibratio

Page 32 - ZERO N

34 • Calibrating Instrumentsfailure and the following screen will appear.In this case you may attempt another bump test, press and release the EDIT E

Page 33

Charging an Instrument in a Calibration Station • 35 17. The results of the calibration will be stored in the SM-2009U’s memory and will be available

Page 34 - BUMP/CAL

Table of ContentsChapter 4: Using the SM-2009U. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25Overview . . . . . . . . . . . . . . .

Page 35 - Calibrating Instruments

36 • Charging an Instrument in a Calibration Stationinstrument by pressing and holding the EDIT ENTER and POWER buttons simultaneously for at least o

Page 36 - 32 • Calibrating Instruments

Calibration and Bump Test Records • 37it, the CHARGE LED will continue to blink yellow while charging is taking place. The SM-2009U will take approxi

Page 37 - CAL

38 • Calibration and Bump Test RecordsNOTE: The USB port on the front of the calibration station cannot be used to connect the SM-2009U to a computer

Page 38 - CAL

Calibration and Bump Test Records • 39NOTE: If you pull out the flash drive while it’s LED is still blinking and the COPY LED is still solid red, the

Page 39 - Charging an Instrument in a

40 • Calibration and Bump Test RecordsBump Testing or Calibrating and Saving Files To a Flash Drive Multiple Times In One DayThe SM-2009U assigns fil

Page 40

Overview • 41Chapter 5: Single Module Data Viewer ProgramOverviewThe Single Module Data Viewer Program is used to view, organize, and print bump test

Page 41 - Flash Drive

42 • Using the Single Module Data Viewer SoftwareNOTE: When you start the Single Module Data Viewer Program for the first time, there will be no data

Page 42

Using the Single Module Data Viewer Software • 43upper right side of the window shows the contents of the item selected in the upper left side, and t

Page 43

44 • Using the Single Module Data Viewer Software 2. Navigate to the location of the files that you want to import into the database. 3. Select the

Page 44

Using the Single Module Data Viewer Software • 45 2. ID View FormatIn ID view format, the data can be organized by one or more of the following items

Page 45

Overview • 1Chapter 1: IntroductionOverviewThis chapter briefly describes the SM-2009U Single Module Calibration Station and the Single Module Data V

Page 46 - Data Viewing Window

46 • Using the Single Module Data Viewer Softwareview the contents below it. Single click on an item to view the contents in the right side of the wi

Page 47

Using the Single Module Data Viewer Software • 47calibration or bump test files will appear in the upper right side of the window.If you are viewing

Page 48 - Organizing the Data

48 • Using the Single Module Data Viewer Softwareinclude the instrument’s serial number, station ID, user ID, bump test or calibration time, test gas

Page 49 - Viewing the Data

Using the Single Module Data Viewer Software • 49Deleting DataYou can delete an instrument, bump test data, or calibration data in the data view wind

Page 50

50 • Spare Parts ListExiting the ProgramTo exit the Single Module Data View Program, do the following: 1. Click the Exit button in the upper right co

Page 51

Spare Parts List • 5181-0158RK-02 Four-gas calibration cylinder, 50% LEL Isobutane/12% O2/50 ppm CO/25 ppm H2S, 58 liter aluminum81-0158RK-04 Four-ga

Page 52

2 • System RequirementsCAUTION: The GX-2009 detects oxygen deficiency and elevated levels of oxygen, combustible gases, carbon monoxide, and hydrogen

Page 53 - Changing the Password

Specifications • 3SpecificationsTable 1: SM-2009U SpecificationsInput Power 12 VDCNOTE: AC Adapter with 100 - 240 VAC, 50/60 Hz, 0.6A input and 12 VD

Page 54 - Spare Parts List

4 • About this ManualAbout this ManualThe SM-2009U Single Module Calibration Stations Operator’s Manual uses the following conventions for notes, cau

Page 55 - Part Number Description

Overview • 5Chapter 2: DescriptionOverviewThis section describes the SM-2009U single module calibration station. It is designed to be used on a table

Comments to this Manuals

No comments