Rki-instruments 65-2435RK User Manual

Browse online or download User Manual for Sensors Rki-instruments 65-2435RK. RKI Instruments 65-2435RK User Manual

  • Download
  • Add to my manuals
  • Print

Summary of Contents

Page 1 - Operator’s Manual

RKI Instruments, Inc.www.rkiinstruments.com65-2435RKHydrogen Compensated CarbonMonoxide TransmitterOperator’s ManualPart Number: 71-0073RKRevision: AR

Page 2 - Product Warranty

10 • 65-2435RK CO TransmitterStart UpThis section describes procedures to start up the CO transmitter and place the transmitter into normal operation

Page 3 - Table of Contents

65-2435RK CO Transmitter • 11MaintenanceThis section describes maintenance procedures. It includes preventive maintenance, troubleshooting, and compo

Page 4 - Specifications

12 • 65-2435RK CO TransmitterPerforming the response test1. Screw the regulator into the calibration cylinder. The sample will begin to flow.2. Allow

Page 5 - Description

65-2435RK CO Transmitter • 13NOTE: Under “normal” circumstances, the transmitter requires calibration once every three months. Some applications may

Page 6 - Junction Box

14 • 65-2435RK CO TransmitterCAUTION: The null potentiometer must be set whenever a sensor is changed or the hydrogen compensation may not work prope

Page 7 - Installation

65-2435RK CO Transmitter • 159. Reconnect the wiring from the controller to the interconnect terminal strip as shown in Table 2 and Figure 3, Wiring

Page 8 - 8 • 65-2435RK CO Transmitter

16 • 65-2435RK CO Transmitter7. Connect the detector leads to the detector terminal strip as shown in Table 4 and Figure 3, Wiring the CO Transmitter

Page 9 - Figure 3

65-2435RK CO Transmitter • 17Setting the Null PotentiometerWhen an old sensor is replaced, the null potentiometer must be adjusted so that the CO tra

Page 10 - Start Up

18 • 65-2435RK CO Transmitter2. Allow the gas to flow for two minutes, then verify a reading of 100 mV (± 2mV). If necessary, use the zero potentiome

Page 11 - Maintenance

65-2435RK CO Transmitter • 19CalibrationThis section describes how to calibrate the CO transmitter. It includes procedures to prepare for calibration

Page 12 - Troubleshooting

2 • 65-2435RK CO TransmitterProduct WarrantyRKI Instruments, Inc., warrants gas alarm equipment sold by us to be free from defects in materials, work

Page 13 - Probable causes

20 • 65-2435RK CO Transmitter2. Allow the gas to flow for two minutes, then verify that the reading matches the response reading (± 2mV) you determin

Page 14 - 14 • 65-2435RK CO Transmitter

65-2435RK CO Transmitter • 21Parts ListTable 5 lists replacement parts and accessories for the CO transmitter.Table 5: Parts ListPart Number Descript

Page 15 - 65-2435RK CO Transmitter • 15

65-2435RK CO Transmitter • 3Table of ContentsOverview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 16 - 16 • 65-2435RK CO Transmitter

4 • 65-2435RK CO TransmitterOverviewThis manual describes the hydrogen compensated carbon monoxide (CO) transmitter. This manual also describes how t

Page 17 - Resetting the Zero Reading

65-2435RK CO Transmitter • 5DescriptionThis section describes the components of the CO transmitter. The transmitter consists of the CO detector, ampl

Page 18 - Returning to Normal Operation

6 • 65-2435RK CO TransmitterCharcoal FilterThe disc-shaped charcoal filter is secured to face of the CO sensor with a rubber boot. The charcoal filte

Page 19 - Calibration

65-2435RK CO Transmitter • 7monitoring site. Use the cover on the front of the junction box to access the interior of the junction box.InstallationTh

Page 20

8 • 65-2435RK CO TransmitterWiring the CO Transmitter to a ControllerWARNING: Always verify that the power source is OFF before you make wiring conne

Page 21 - Parts List

65-2435RK CO Transmitter • 99. Connect the wires to the to the applicable controller transmitter terminal strip as shown in Figure 3.Figure 3: Wiring