www.rkiinstruments.com65-2424RKHydrogen Sulfide TransmitterOperator’s ManualPart Number: 71-0062RKRevision: AReleased: 2/16/11

Page 2 - 65-2424RK H

6 • 65-2424RK H2S Transmitter 9. If shielded cable is used, connect the cable’s drain wire to an available chassis (earth) ground at the controller.

Page 3 - Product Warranty

65-2424RK H2S Transmitter • 7reading is 100 mV (±2 mV).11. Turn the regulator on/off knob clockwise to close it.12. Remove the calibration cup from t

Page 4 - Table of Contents

8 • 65-2424RK H2S Transmitter 7. Use the following formula to determine the correct test points output for the test sample.Output (mV) = (calibrating

Page 5 - Specifications

65-2424RK H2S Transmitter • 9Recommended action• Verify that the transmitter wiring is correct and secure.• Calibrate the transmitter.• If the fail c

Page 6 - Description

10 • 65-2424RK H2S Transmitter 7. Secure the cover to the enclosure.8. Turn on or plug in incoming power at the power source end.9. Turn on the contr

Page 7 - Enclosure

65-2424RK H2S Transmitter • 1115. Calibrate the H2S transmitter as described in the Calibration section of this insert.Calibration FrequencyAlthough

Page 8 - Installation

12 • 65-2424RK H2S Transmitter Setting the Zero ReadingNOTE If you can verify that the H2S transmitter is in a fresh air environment, you do not need

Page 9 - S Transmitter to a Controller

65-2424RK H2S Transmitter • 135. Store the components of the calibration kit in a safe and convenient place.Parts ListTable 6 lists replacement parts

Page 10 - Start Up

65-2424RK H2S Transmitter WARNINGRead and understand this instruction manual before operating detector. Improper use of the detector could result in

Page 11 - Maintenance

65-2424RK H2S TransmitterProduct WarrantyRKI Instruments, Inc. warranties gas alarm equipment sold by us to be free from defects in materials, workma

Page 12 - Troubleshooting

65-2424RK H2S Transmitter Table of ContentsOverview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .

Page 13 - S Transmitter

65-2424RK H2S Transmitter • 1OverviewThis manual describes the 65-2424RK hydrogen sulfide (H2S) transmitter. This manual also describes how to instal

Page 14

2 • 65-2424RK H2S Transmitter DescriptionThis section describes the components of the H2S transmitter. The transmitter consists of the H2S sensor, se

Page 15 - Calibration Frequency

65-2424RK H2S Transmitter • 3Interconnect Terminal StripThe interconnect terminal strip is the four-point terminal strip on the lower left side of th

Page 16 - Returning to Normal Operation

4 • 65-2424RK H2S Transmitter InstallationThis section describes procedures to mount the H2S transmitter in the monitoring environment and wire the t

Page 17 - Parts List

65-2424RK H2S Transmitter • 5Wiring the H2S Transmitter to a ControllerWARNING: Always verify that the power source is off before you make wiring con