www.rkiinstruments.com35-3010RK-05 Sample Draw DetectorPart Number: 71-0210RKRevision: BReleased: 9/10/12
6 • 35-3010RK-05 Sample-Draw DetectorMain Circuit BoardFigure 3: Main Circuit BoardThe main circuit board includes the interconnect terminal strip,
35-3010RK-05 Sample-Draw Detector • 7Amp 1 and Amp 2 CircuitsThese circuits are located to the left of the sensor/transmitter terminal strip. They ea
8 • 35-3010RK-05 Sample-Draw DetectorInstallationThis section describes procedures to mount the sample draw gas detector in the monitoring environmen
35-3010RK-05 Sample-Draw Detector • 96. Position the sample draw housing on a vertical surface at eye level (4 1/2 to 5 feet from the floor).7. Inser
10 • 35-3010RK-05 Sample-Draw DetectorFigure 5: External (Field) Wiring, Sample Draw DetectorInterconnect Terminal Strip(G)24 VDC -COController Tran
35-3010RK-05 Sample-Draw Detector • 11Figure 6: Internal (Factory) Wiring, Sample Draw DetectorPressure Switch Factory WiredPumpFactoryWiredH2SCO Se
12 • 35-3010RK-05 Sample-Draw DetectorStart UpThis section describes procedures to start up the sample draw detector and place the sample draw detect
35-3010RK-05 Sample-Draw Detector • 13the controller. See the controller operator’s manual for directions.6. Remove the voltmeter from the test point
14 • 35-3010RK-05 Sample-Draw Detector1. Verify that the controller is reading 20.9% for the oxygen channel and 0 ppm for the CO channel.If the readi
35-3010RK-05 Sample-Draw Detector • 153. Verify that the sensor wiring is correct and secure. “Wiring the Sample Draw Detector” on page 9 describes d
35-3010RK-05 Sample-Draw DetectorWARNINGRead and understand this instruction manual before operating detector. Improper use of the detector could res
16 • 35-3010RK-05 Sample-Draw Detector6. Place the sensor in the oxygen sensor cavity, then position the retaining plate on the two standoffs.7. Secu
35-3010RK-05 Sample-Draw Detector • 17If the sample draw detector goes into low flow alarm before you can adjust the flow down to 0.6 SCFH, adjust th
18 • 35-3010RK-05 Sample-Draw Detector2. Connect the calibration kit sample tubing to the regulator.3. Follow the directions in the controller’s oper
35-3010RK-05 Sample-Draw Detector • 19Preparing for CalibrationCAUTION: This procedure may cause alarms at the controller. Take appropriate action to
20 • 35-3010RK-05 Sample-Draw DetectorNOTE: For convenience, leave the regulator connected to the sample tubing.Returning to Normal Operation1. Remov
35-3010RK-05 Sample-Draw DetectorProduct WarrantyRKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in materials,
35-3010RK-05 Sample-Draw DetectorTable of ContentsOverview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
35-3010RK-05 Sample-Draw Detector • 1OverviewThis manual describes the 35-3010RK-05 sample draw detector. This manual also describes how to install,
2 • 35-3010RK-05 Sample-Draw DetectorDescriptionThis section describes the components of the 35-3010RK-05 sample draw detector head. The sample draw
35-3010RK-05 Sample-Draw Detector • 3An aluminum subpanel is mounted to the interior of the housing. The sample draw detector’s internal components a
4 • 35-3010RK-05 Sample-Draw DetectorCharcoal FilterThe charcoal filter is located in the top left corner of the enclosure. It is held in place by a
35-3010RK-05 Sample-Draw Detector • 5Flow BlockThe flow block is located in the lower right corner of the sample draw detector. All the sensors are i