www.rkiinstruments.com61-1004RK/61-0191RKCO2 DetectorOperator’s ManualPart Number: 71-0145RKRevision: BReleased: 7/18/14
6 • 61-1004RK/61-0191RK Carbon Dioxide DetectorFigure 3: Wiring the CO2 Detector to a Controller10. If using shielded cable, connect the cable’s dra
61-1004RK/61-0191RK Carbon Dioxide Detector • 7Start UpThis section describes procedures to start up the CO2 detector and place the detector into nor
8 • 61-1004RK/61-0191RK Carbon Dioxide DetectorMaintenanceThis section describes maintenance procedures. It includes preventive maintenance, troubles
61-1004RK/61-0191RK Carbon Dioxide Detector • 9NOTE: If the reading is not within ± 20% of the gas concentration, calibrate the detector as described
10 • 61-1004RK/61-0191RK Carbon Dioxide DetectorReplacing the CO2 Detector1. Turn off the controller.2. Turn off power to the controller.3. If the de
61-1004RK/61-0191RK Carbon Dioxide Detector • 11then more frequent calibration than every 6 months may be necessary.CalibrationThis section describes
12 • 61-1004RK/61-0191RK Carbon Dioxide Detector4. Set the response reading according to the controller’s operator’s manual.5. Unscrew the CO2 cylind
61-1004RK/61-0191RK Carbon Dioxide Detector • 13Parts ListTable 3 lists replacement parts and accessories for the 61-1004RK and 61-0191RK CO2 detecto
61-1004RK/61-0191RK Carbon Dioxide DetectorWARNINGRead and understand this instruction manual before operating detector. Improper use of the detector
61-1004RK/61-0191RK Carbon Dioxide DetectorProduct WarrantyRKI Instruments, Inc. warrants gas alarm equipment sold by us to be free from defects in m
61-1004RK/61-0191RK Carbon Dioxide DetectorTable of ContentsOverview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
61-1004RK/61-0191RK Carbon Dioxide Detector • 1OverviewThis manual describes the 61-1004RK CO2 (carbon dioxide) detector. This manual also describes
2 • 61-1004RK/61-0191RK Carbon Dioxide DetectorDescriptionThe 61-1004RK CO2 detector has four versions, the 61-1004RK-02 which has a detection range
61-1004RK/61-0191RK Carbon Dioxide Detector • 3The output of the CO2 detector mimics the output of RKI’s standard catalytic LEL detectors. For this r
4 • 61-1004RK/61-0191RK Carbon Dioxide DetectorInstallationThis section describes procedures to mount the CO2 detector in the monitoring environment
61-1004RK/61-0191RK Carbon Dioxide Detector • 5Wiring the CO2 Detector to a ControllerWARNING: Always verify that the power to the controller is off