73 Replacing the HC and CO Filters GX-2003 Operator’s Manual
2. Insert the battery pack into the charging station as shown in
Figure 8 below so that the metal contacts on the ends of the battery
pack come into contact with the metal contacts in the charging
station. When proper contact has been made, the red LED on the
charging station will turn on.
Figure 8: Putting the Battery Pack into the Charging Station
3. When a full charge has been reached, approximately 90 minutes,
the red LED on the charging station will turn off. Remove the
battery pack from the charging station by lifting up on the tab with
the arrowhead on it and unplug the charging station’s power cord
from the AC outlet.
Replacing the HC and CO Filters
1. Verify that the GX-2003 is off.
2. Use a coin to unscrew the plastic filter retainer from the flow
chamber on the back of the GX-2003. The HC filter is in the
position closer to the exhaust and has a yellow label around its
position. The CO filter is below the HC filter and has a red label
around its position.
3. If the filter does not come out with the filter retainer, carefully
remove it from the flow chamber.
4. Discard the old filter and snap the colored side (yellow for HC and
red for CO) of the new filter into the filter retainer.
Red Charging
LED
Side View
Top View
Polarity Indexing Marks
Pull Up On This Tab
To Remove Pack