RKI Instruments GX-2003 Operator's Manual Page 39

  • Download
  • Add to my manuals
  • Print
  • Page
    / 94
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 38
GX-2003 Operator’s Manual Calibration Mode • 32
Calibration Mode
Calibration mode allows you to calibrate the GX-2003’s sensor
response. It also includes a menu item which allows you to adjust the
LCD contrast.
CAUTION:BEFORE EACH DAY’S USAGE SENSITIVITY IN THE
% LEL RANGE MUST BE TESTED ON A KNOWN
CONCENTRATION OF THE COMBUSTIBLE TARGET
GAS, METHANE, EQUIVALENT TO 25 - 50% OF FULL
SCALE CONCENTRATION (The % LEL full scale is
100% LEL). ACCURACY MUST BE WITHIN -0 to + 20%
OF ACTUAL. ACCURACY MAY BE CORRECTED BY
FOLLOWING THE CALIBRATION INSTRUCTIONS FOR
THE COMBUSTIBLE CHANNEL BELOW.
If the combustible channel passes the above response test and does
not require calibration, the unit should still be calibrated periodically.
The optimum frequency of calibration depends heavily on how the
GX-2003 is used. For example, instruments used daily may need to be
calibrated weekly or monthly, while instruments that are used only a
few times a year may need to be calibrated before each use. Typical
calibration frequencies range from monthly to quarterly. Make sure to
perform the combustible channel response test as described above
and make sure to develop a calibration schedule tailored to your
application that takes this test and required calibration resulting from
this test into account.
You can program the GX-2003 to notify you when it is due for
calibration. See “Updating the Calibration Time Remaining Setting”
on page 59.
Calibration Supplies and Equipment
To calibrate the GX-2003, you will need:
Known calibrating samples of the gases being detected. The
combustible and toxic gas samples should have concentrations in
approximately the middle of the range of detection. An oxygen-
free source, such as 100% nitrogen is recommended for setting the
oxygen zero.
A demand-flow regulator to provide adequate sample gas flow
Non-absorbent tubing
Page view 38
1 2 ... 34 35 36 37 38 39 40 41 42 43 44 ... 93 94

Comments to this Manuals

No comments