Rki-instruments 35-3001-02 User Manual Page 16

  • Download
  • Add to my manuals
  • Print
  • Page
    / 21
  • Table of contents
  • TROUBLESHOOTING
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 15
12 • 35-3001-02 Oxygen Sample-Draw Detector
Maintenance
This section describes maintenance procedures. It includes preventive maintenance procedures.
This section also includes procedures to troubleshoot the sample-draw detector, replace
components of the sample-draw detector, and adjust the low flow setting.
Preventive Maintenance
This section describes a preventive maintenance schedule to ensure the optimum performance of
the sample-draw detector. It includes daily, monthly, and quarterly procedures.
Daily Visual Checks
1. Verify that the Pilot LED is on.
2. Verify that the flowmeter indicates a flow rate of approximately 1.0 SCFH. If necessary use the
sensor flow control valve to adjust the flow rate to 1.0 SCFH. Turn the valve’s knob
counterclockwise to increase the flow and clockwise to decrease the flow.
3. Verify a display reading of 20.9% at the controller. Investigate significant changes in the
display reading.
Monthly Response Test
This procedure describes a test to verify that the sample-draw detector responds properly to the
target gas.
NOTE: To reduce the response time of this test, use a short incoming sample line. If the sample-
draw detectors sample line is long, connect a shorter line for this test. Make sure you
reconnect the sample line after you complete this procedure.
NOTE: Performing a response test on the sample-draw detector may cause alarms. Be sure to put
the controller into its calibration program or disable external alarms before performing
this test.
Preparing for the response test
NOTE: This procedure describes the RKI calibration kit that includes a demand flow regulator.
1. Verify that the display reading at the controller is 20.9%.
If the display reading is not 20.9%, set the fresh air reading as described in “Calibration” on
page 16, then continue this procedure.
Performing the response test
1. Exhale into the sample-draw detectors inlet fitting.
2. Stop exhaling into the sample line, then verify that the oxygen reading at the controller
decreased.
3. If the reading does not decrease, calibrate the sample draw detector as described in
“Calibration” on page 16.
4. Reconnect the sample tubing to the inlet fitting.
Quarterly Calibration
Calibrate the sample-draw detector as described in “Calibration” on page 16.
Page view 15
1 2 ... 11 12 13 14 15 16 17 18 19 20 21

Comments to this Manuals

No comments