Page view 56
Viewing, Printing, Exporting, and Deleting Data in the Data Window • 53
• In table format, the log times are shown along with the
peak (minimum for oxygen) gas readings for the previous
five seconds.
• The gas readings at the time of the alarm event around
which the logged data are centered are highlighted in red
and are the instantaneous readings at that time.
• The catalytic combustible channel can be recorded in
%LEL, ppm, or %volume units depending on the
instrument setting. When viewing the alarm trend data in
table format, the units for the catalytic LEL channel are
shown in %LEL. The displayed units can be changed to
ppm or %volume using the selection box in the bottom left
portion of the screen.