35-3000RK-10 PPM Hexane Sample Draw Detector 11
NOTE: The sample draw detector is shipped with the mounting feet “tucked under” the
housing to protect the mounting feet during shipment.
3. Slightly loosen the screw that secures one of the mounting feet to the housing, then
rotate the mounting foot 180 degrees (see Figure 3).
4. Tighten the screw that secures the mounting foot to the housing.
5. Repeat steps 3 and 4 for the remaining three mounting feet.
6. Position the sample draw housing on a vertical surface at eye level (4 1/2 to 5 feet
from the floor).
7. Insert 1/4 in. or 5/16 screws through the slots in the mounting feet to secure the
housing to the mounting surface.
Connecting the Sample Lines to the Sample Draw Detector
It is important to use the appropriate tubing material for the 0 - 500 ppm hexane sample
draw detector. To avoid the possibility of losing sample in the inlet tubing, use either
stainless steel, rigid polypropylene, or flexible polyurethane tubing.
1. Attach 1/4 inch O.D. stainless steel, polypropylene, or polyurethane sample tubing to
the inlet fitting.
CAUTION: If you use flexible sample tubing (polyurethane for example), use an appropriate
metal insert to seal the connection between the tubing and the INLET fitting. See
“Parts List” on page 26 for an example of an appropriate metal insert.
2. Place the opposite end of the tubing in the sampling area.
CAUTION: Avoid loops or slumps in the incoming sample line. To reduce response time, keep the
incoming sample line as short as possible.
3. Attach rigid sample tubing or flexible tubing with an appropriate insert to the exhaust
fitting.
4. Route the opposite end of the tubing to an open area where the sample can safely
disperse.
Wiring the Sample Draw Detector
WARNING: Always verify that all power is off before making any wiring connections
1. Turn off the controller.
2. Turn off power to the controller.
3. Unlatch and open the housing door of the sample draw detector.
4. Guide a three-conductor, shielded cable or three wires in conduit through one of the
conduit hubs at the bottom of the sample draw detector housing.
5. Connect the cable or wires in conduit to the sample draw detector’s interconnect
terminal strip as shown in Figure 4.
6. Close and latch the housing door of the sample-draw detector.