Rki-instruments GX-2009 User Manual Page 18

  • Download
  • Add to my manuals
  • Print
  • Page
    / 54
  • Table of contents
  • BOOKMARKS
  • Rated. / 5. Based on customer reviews
Page view 17
18 • Connecting a GX-2009 to the User Setup Program
5. There are two additional exceptions to the sequence
described in step 3 above if the unit includes the bump test
function and if Bump Limit Display is set to On. See “Detail
Settings” on page 44 for a description of the Bump Limit
Display and Bump Limit Check parameters.
When Bump Limit Display is set to On and Bump Limit
Check is set to Can’t Use, if the GX-2009 is due for bump
testing the instrument will show a bump test expired failure
screen and the buzzer and LED arrays will pulse for a few
seconds and before connecting to the program. The
failure screen will remain on the display while the unit is
connected to the program. You can still use the User
Setup Program with an instrument if this failure indication
occurs.
When Bump Limit Display is set to On and Bump Limit
Check is set to Confirm to use, if the GX-2009 is due for
bump testing the unit will show the following screen and
beep and flash the LED arrays for a few seconds
indicating that bump testing is due.
When the beeping and flashing stops, press and release
the POWER MODE button on the GX-2009 to continue.
The unit will now connect to the program and the PC
TRANSMIT screen will be displayed.
You can now retrieve the connected instrument’s configuration
information using either the Instrument Information or the
Information Get & Save command button as described in
“Instrument Information” on page 20 and “Information Get &
Save” on page 23. You must retrieve the connected instrument’s
configuration information before you can perform some
operations such as saving the instrument’s parameter
FAIL
B--LIMIT
tESt
B--LIMIT
Page view 17
1 2 ... 13 14 15 16 17 18 19 20 21 22 23 ... 53 54

Comments to this Manuals

No comments