Sample Draw Combustible Gas Detector • 1 35-3000RKA-LEL Sample Draw Combustible Gas Detector Specifications Table 1 lists specifications for the samp
10 • Sample Draw Combustible Gas Detector Start Up This section describes procedures to start up the sample draw combustible gas detector and place
Sample Draw Combustible Gas Detector • 11 Maintenance This section describes maintenance procedures. It includes preventive maintenance procedures.
12 • Sample Draw Combustible Gas Detector 8. When the bag is full, unscrew the fixed flow regulator form the cylinder.9. Close the clamp, then discon
Sample Draw Combustible Gas Detector • 13 Slow or no response/difficult or unable to calibrate Symptoms • The detector responds slowly or does not r
14 • Sample Draw Combustible Gas DetectorReplacing Components of the Combustible GasSample Draw DetectorThis section includes procedures to replace
Sample Draw Combustible Gas Detector • 15Adjusting the Low Flow SettingThe factory-set low flow setting is 0.6 SCFH (±0.2). To adjust the low flow sett
16 • Sample Draw Combustible Gas DetectorCalibrationThis section describes how to calibrate the sample draw combustible gas detector. It includes pro
Sample Draw Combustible Gas Detector • 175. Connect the tubing from the gas collection bag to the fixed flow regulator, then open the clamp.6. Screw th
18 • Sample Draw Combustible Gas DetectorParts ListTable 4 lists replacement parts and accessories for the sample draw combustible gas detector.Table
2 • Sample Draw Combustible Gas Detector Description This section describes the components of the sample draw combustible gas detector. The sample
Sample Draw Combustible Gas Detector • 3 Flow System The sample draw detector’s flow system consists of the INLET fitting, filter, pump, flowmeter, byp
4 • Sample Draw Combustible Gas Detector Status lights Two status lights are above the flowmeter. They are also visible through the window in the ho
Sample Draw Combustible Gas Detector • 5 Connector The top of the sensor includes five pins that plug into the socket connector. This connector allo
6 • Sample Draw Combustible Gas Detector Pump terminal strip The pump terminal strip is the four-point terminal strip near the top edge of the circ
Sample Draw Combustible Gas Detector • 7 Installation This section describes procedures to mount the sample draw combustible gas detector in the mo
8 • Sample Draw Combustible Gas Detector 2. Close and latch the housing door. NOTE: The sample draw detector is shipped with the mounting feet “tuc
Sample Draw Combustible Gas Detector • 9 5. Route the cable or wires in conduit leading from the sample draw detector to the recording or monitorin
Comments to this Manuals